Echocardiographic evaluation of prosthetic heart valves

Nelson B Schiller, MD
Elyse Foster, MD
Nov 4, 1999

Delineation of normal prosthetic valve function is usually possible with transthoracic echocardiography. Despite the prominent acoustic shadowing that accompanies mechanical prostheses, a Doppler signal that demonstrates normal transprosthetic flow velocity and flow duration is usually sufficient to exclude a stenotic valve. Exclusion of an incompetent valve is often more difficult, especially for mechanical prostheses in the mitral position. Transthoracic Doppler echocardiography provides important hemodynamic information with regard to prosthetic valve pressure gradients and is extremely useful in long-term follow-up. The "normal gradient" across a prosthetic valve depends upon the type, size, and position of the prosthesis as well as the cardiac output; guidelines are available for the acceptable range of Doppler gradients encountered in properly functioning valves [1,2].

Acoustic shadowing caused by the prosthetic material may limit transthoracic visualization of prosthetic leaflets, vegetations, and thrombi. In addition, while prosthetic aortic valve regurgitation is usually well visualized on surface color Doppler imaging, prosthetic mitral regurgitation is frequently undetectable [3]. As a result, transesophageal echocardiography (TEE) is the imaging method of choice when the transthoracic echocardiogram is technically inadequate or when there are borderline findings on the transthoracic echocardiogram in a patient in whom there is a strong clinical suspicion of malfunction [4].

Transthoracic echocardiography images anteriorly, and thus is usually sufficient for evaluation of aortic prostheses. However, since the atria are situated in the far field where resolution is low, they are shadowed by a mechanical mitral prosthesis. Since TEE images are taken from a position posterior to the heart, the atrial side of the mitral and tricuspid valves can be seen. On the other hand, transthoracic echocardiography demonstrates the ventricular side of the valve that lies in its near field, whereas TEE does not. Both methods may have difficulty with the evaluation of aortic prostheses, and the presence of mitral and aortic devices in the same patient may make the task of the echocardiographer more difficult. In most cases, however, both echocardiographic methods are complementary.

The physician evaluating a prosthetic device with TEE should be aware of the range of abnormalities that are possible in these devices and should match those possibilities to the patient's presentation. Complications of prosthetic valves detected by TEE include:

  •  Perivalvular leak
  •  Endocarditis
  •  Extrinsic interference of function (pannus, thrombus, vegetation) resulting in obstruction or regurgitation
  •  Ball variance
  •  Strut fracture and component escape
  •  Leaflet tears of bioprosthesis
  •  Leaflet calcification/stenosis of bioprosthesis

The role of the TEE for evaluating the abnormalities associated with prosthetic valves will be reviewed here. The major complications encountered with these valves and the role of echocardiography in the evaluation of infective endocarditis are discussed fully elsewhere. (See "Complications of prosthetic heart valves" and see "Role of echocardiography in infective endocarditis-I").

PERIVALVULAR REGURGITATION ! Perivalvular regurgitation can arise from broken or dehisced sutures, from a poorly seated ring, or from endocarditis (dehiscence). Hemolysis is a common, but generally underappreciated, complication of these leaks, especially when they occur with a mitral valve prosthesis [5,6]. A TEE is required to detect perivalvular leakage and should routinely be performed when a patient with a prosthetic valve presents with severe anemia. Severe hemolysis can also occur after mitral valve repair, when regurgitation develops around the annuloplasty ring. (See "Extrinsic nonimmune hemolytic anemia due to mechanical damage: Fragmentation hemolysis and hypersplenism", section on Hemodynamic turbulence).

To recognize a perivalvular leak, TEE must be performed with a high color frame rate and middle range Nyquist limits (35 to 50 cm/sec) in several views from several angles outside the sewing ring [7]. There should be a careful search for periprosthetic leaks around as much of the valve circumference as possible and an attempt to define the extent of the regurgitation once it is identified. The origin of a periprosthetic leak may appear deceivingly narrow when it is related to disruption of a limited number of sutures. In some cases, these jets are seen inadvertently when one examines other structures, such as the interatrial septum.

The most severe form of perivalvular regurgitation is seen when there is dehiscence of a substantial portion of the sewing ring [8]. In this setting, there is severe regurgitation, which may be so torrential that the regurgitant flow is almost laminar. As in angiography, the sewing ring on echocardiography is seen to rock with each cardiac cycle; mobile echo densities representing the liberated suture material often can be visualized. Both the laminar flow and the rocking motion of the ring may elude the inexperienced echocardiographer, delaying recognition of this life threatening condition.

PROSTHETIC VALVE OBSTRUCTION ! When, in a newly symptomatic patient, there is an unexpected rise in transprosthetic gradient from a baseline determination or from established normal values for valves of that type and size, the possibility of progressive obstruction should be considered. Causes of obstruction include pannus ingrowth, thrombus, and vegetation. Clinical clues to this possibility include the age of the valve and the adequacy of anticoagulation. In a bioprosthesis or heterograft, the leaflets themselves may become calcified and immobile.

Once there is a high suspicion of obstruction, TEE should be strongly considered for both mechanical and bioprosthetic valves.

  •  In the case of suspected aortic pannus, the distal end of the left ventricular outflow tract should be examined both with imaging and with color flow. Pannus tends to lie close to the valve ring and can be easily overlooked. There can be a prevalve jet that suggests pannus, which, on further searching with a variety of frequencies, angles, and gain settings, will be detected.

  •  In the mitral position, the same procedure should be followed. Finding a high grade of spontaneous contrast in the left atrium, with or without thrombi, or finding thrombus around the sewing ring in the setting of adequate anticoagulation should heighten suspicion of pannus formation. Thin fibrillar strands are also encountered on the mitral annulus and the left ventricular outflow tract. These structures are brightly reflective and highly mobile and may or may not be associated with a pathologic process.

When a thrombus is present in the vicinity of the sewing ring and when obstruction is clinically plausible and suggested by Doppler imaging, consideration should be given to the use of thrombolysis or surgery [9]. (See "Complications of prosthetic heart valves").

Distinction between thrombus and pannus ! The most common etiology for prosthetic valve obstruction is thrombus formation; pannus formation due to fibrous tissue ingrowth is less common. In one surgical study of 112 obstructed mechanical valves, pannus formation was the underlying cause in 11 percent of valves, pannus formation in combination with thrombus was present in 12 percent, while thrombus alone was the etiology in the remaining cases [10]. Though associated with substantial morbidity, thrombolytic therapy is often an effective alternative to surgery for treatment of thrombus. Therefore, it is important to distinguish between these two causes. (See "Complications of prosthetic heart valves", section on Pannus formation).

Echocardiographic differentiation of pannus and thrombus may be difficult. In general:

  •  Thrombus tends to be mobile, somewhat less echo-dense, and associated with spontaneous contrast

  •  Pannus is highly echogenic, consistent with its fibrous composition, and is usually firmly fixed to the valve apparatus

Color flow aliasing with proximal acceleration of the flow jet in the vicinity of the mass may aid in the identification of pannus. In one study of 23 patients who presented with 24 obstructed valves, clinical, transthoracic, and TEE data were compared to pathology upon surgery in order to determine the clinical and echocardiographic characteristics that differentiate thrombus from pannus formation [11]. Thrombus was established in 14  valves and pannus in 10. Pannus formation was more common in the aortic position. From the standpoint of echocardiography, thrombus was more likely to be associated with soft ultrasound density (92 versus 29 percent).

PROSTHETIC VALVE REGURGITATION ! Physiologic regurgitation, the so-called seating puff of angiography, is universally encountered and dependent in degree on the type of prosthesis used. However, severe regurgitation may result from bioprosthetic valve degeneration , mechanical valve pannus, thrombus, or vegetation.

Physiologic regurgitation ! All mechanical valves exhibit some degree of obligatory regurgitation of up to 15 mL of blood [12]. Seating regurgitation or "closure backflow" appears only briefly and is due to retrograde volume displacement as the valve leaflets close. This type of regurgitation is detected by highly sensitive color flow Doppler imaging on TEE. In addition, a certain amount of more prolonged "leakage backflow" regurgitation occurs after the valve closes [13].

In an evaluation of 136 mechanical prosthetic valves, TEE color Doppler imaging revealed regurgitant jets in 95 percent of the mitral prostheses and 44 percent of the aortic prostheses [14]. In contrast transthoracic echocardiography documented regurgitant jets in only 28 and 29 percent, respectively. The lower incidence of detected regurgitation in the aortic position may be explained by the shadowing of the left ventricular outflow tract in TEE imaging [15].

Normally functioning mechanical valves, such as the bileaflet St. Jude, usually have two to four centrally directed regurgitant jets. Features associated with these jets include a low intensity and only minimal penetration into the atrium, generally less than 3 cm [13,16,17]. The monodisc Medtronic-Hall valve has two jets, one of which is prominent and longer [17,18]. Normally functioning heterografts are less likely to have these small regurgitant signals; when mild regurgitation is present, there is usually one central jet [14,19].

Pathologic regurgitation ! Pathologic jets tend to be high velocity, central, intense, broad, and highly aliased; they tend to be eccentric, hugging the wall of the atrium (Coanda effect), penetrating to its roof, continuing around its perimeter and terminating where they began. The offending mass or tissue can usually be seen, but the actual locus of obstruction causing the regurgitation may not be visible.

In contrast to mechanical valves, bioprostheses with leaflet degeneration may exhibit central pathologic regurgitation that is broad-based when severe. The bioprosthetic leaflets can usually be seen and are often thickened and calcified with possible perforation or disruption leading to a flail leaflet. Forward flow velocities measured by spectral Doppler may be increased due to the larger flow volume associated with regurgitation as well as concurrent obstruction.

REGURGITATION DUE TO STRUT FRACTURE AND DISK LIBERATIONS ! There has been considerable concern about large Bjork-Shiley monodisc valves because of their low, but significant, incidence of strut fracture, leading to liberation of the disk and rapid hemodynamic compromise. Fortunately, this is very uncommon. When it does occur, clinical deterioration is often so rapid that diagnostic TEE cannot be performed and the patient is frequently taken from the emergency room to surgery based on clinical suspicion.

1. Reisner, SA, Meltzer, RS. Normal values of prosthetic valve Doppler echocardiographic parameters: A review. J Am Soc Echocardiogr 1988; 1:201.
2. Khandheria, BK. Transesophageal echocardiography in the evaluation of prosthetic valves. Cardiol Clin 1993; 11:427.
3. Cohen, GI, Davison, MB, Klein, AL, et al. A comparison of flow convergence with other transthoracic echocardiographic indexes of prosthetic mitral regurgitation. J Am Soc Echocardiogr 1992; 5:620.
4. Khandheria, BK. Transesophageal echocardiography in the evaluation of prosthetic valves. Cardiol Clin 1993; 11:427.
5. Dilip, KA, Vachaspathy, P, Clarke, B, et al. Haemolysis following mitral valve repair. J Cardiovasc Surg (Torino) 1992; 33:568.
6. Smith, RE, Berg, D. Occult paravalvular leak in a clinically normal St. Jude's mitral valve presenting with life-threatening microangiopathic hemolytic anemia. J Cardiovasc Surg (Torino) 1991; 32:56.
7. Groundstroem, K, Rittoo, D, Hoffman, P, et al. Additional value of biplane transoesophageal imaging in assessment of mitral valve prostheses. Br Heart J 1993; 70:259.
8. Akamatsu, S, Ueda, N, Terazawa, E, et al. Mitral prosthetic dehiscence with laminar regurgitant flow signals assessed by transesophageal echocardiography. Chest 1993; 104:1911.
9. Lengyel, M, Fuster, V, Keltai, M, et al. Guidelines for management of left-sided prosthetic valve thrombosis: A role for thrombolytic therapy. J Am Coll Cardiol 1997; 30:1521.
10. Deviri, E, Sareli, P, Wisenbaugh, T, et al. Obstruction of mechanical heart prosthesis: Clinical aspects and surgical management. J Am Coll Cardiol 1991; 17:646.
11. Barbetseas, J, Nagueh, SF, Pitsavos, C, et al. Differentiating thrombus from pannus formation in obstructed mechanical prosthetic valves: An evaluation of clinical, transthoracic and transesophageal echocardiographic parameters. J Am Coll Cardiol 1998; 32:1410.
12. Baumgartner, H, Czer, L, DeRobertis, M, et al. Normal regurgitation in mechanical valve prostheses: Impact on Doppler Studies. American College of Cardiology Learning Center Highlights 1992; 8:1.
13. Baumgartner, H, Khan, S, DeRobertis, M, et al. Color Doppler regurgitant characteristics of normal mechanical mitral valve prostheses in vitro. Circulation 1992; 85:323.
14. Mohr-Kahaly, S, Kupferwasser, I, Erbel, R, et al. Regurgitant flow in apparently normal valve prostheses: improved detection and semiquantitative analysis by transesophageal two- dimensional color-coded Doppler echocardiography. J Am Soc Echocardiogr 1990; 3:187.
15. Karalis, DG, Chandrasekaran, K, Ross, JJ Jr, et al. Single-plane transesophageal echocardiography for assessing function of mechanical or bioprosthetic valves in the aortic valve position. Am J Cardiol 1992; 69:1310.
16. Lange, HW, Olson, JD, Pedersen, WR, et al. Transesophageal color Doppler echocardiography of the normal St. Jude Medical mitral valve prosthesis. Am Heart J 1991; 122:489.
17. Flachskampf, FA, O'Shea, JP, Griffin, BP, et al. Patterns of normal transvalvular regurgitation in mechanical valve prostheses. J Am Coll Cardiol 1991; 18:1493.
18. Hixson, CS, Smith, MD, Mattson, MD, et al. Comparison of transesophageal color flow Doppler imaging of normal mitral regurgitant jets in St. Jude Medical and Medtronic Hall cardiac prostheses. J Am Soc Echocardiogr 1992; 5:57.
19. Chen, YT, Kan, MN, Chen, JS, et al. Detection of prosthetic mitral valve leak: A comparative study using transesophageal echocardiography, transthoracic echocardiography, and auscultation. J Clin Ultrasound 1990; 18:557.