Placement and management of arterial catheters
Arterial cannulation, a commonly performed procedure in the intensive care
unit, simplifies the monitoring and management of many critically ill
patients. Although the procedure has been performed for over 60 years, there
have been significant technological advances in the monitoring system. These
include internal calibration systems, better filtering capabilities, and
visually pleasing color displays. Transparent and disposable solid-state
systems have also replaced troublesome wire elements in transducers.
An arterial line may be used for the following purposes:
• Monitoring of blood pressure in the setting of
hypotension or hypertension, particularly if the problem is acute or the blood
pressure is labile
• Monitoring response to vasoactive drugs
• Obtaining frequent arterial specimens for blood
gas determination
• Obtaining frequent blood specimens in lieu of
venipuncture
• Emergent administration of medications when
venous access cannot be obtained
• Intraaortic balloon pump use
TECHNIQUE ! The required equipment for
arterial cannulation is listed in Table 1 (
show
table 1). Commonly used sites are the radial, femoral, axillary, dorsalis
pedis, brachial, and, in children, temporal and umbilical arteries. Although
each site has unique complications, currently available data do not support
the superiority of any particular site [
1].
The traditional bias against femoral artery cannulation is contradicted by
multiple studies demonstrating a complication rate no higher, and in some
studies lower, than radial artery cannulation [
2].
One study comparing medical and surgical intensive care units (ICU) concluded
that the preferred site for arterial cannulation in medical ICUs is femoral
and in surgical ICUs is radial. In this study, the site of arterial catheter
placement, and the timing and number of catheter/site changes made no
significant difference in terms of complications [
3].
The anatomy and landmarks of the radial, brachial, femoral, and dorsal pedis
arteries are shown in Figures 1 through 4, respectively (
show
figure 1A-1B,
show
figure 2A-2B,
show
figure 3A-3B, and
show
figure 4A-4B).
When placing a radial catheter, the radial artery is palpated between the
distal radius and the tendon of the flexor carpi radialis; it can usually be
more easily accessed with the wrist extended. The modified Allen's test should
be performed to demonstrate collateral flow through the superficial palmar
arch prior to cannulation (
show
figure 5).
Direct puncture or guidewire approaches may be employed for arterial
cannulation. One study of 69 critically ill patients reported that guidewire
techniques yielded a lower failure rate, fewer passes, and more rapid
cannulation than the direct-puncture method [
4].
The guidewire technique is depicted in Figure 6 (
show
figure 6). We recommend the guidewire technique during training years and
also if there are difficulties encountered with the direct puncture technique.
Successful cannulation is more likely in males (because of larger vessel
size), and in patients with bounding pulses [
5].
Devices which incorporate a disposable ultrasound transducer within the lumen
of a thin-wall vascular needle (Smart Needle) are used infrequently, but may
be helpful in certain situations. As an example, hypotensive patients with
nonpalpable pulses may be difficult to cannulate unless instruments or
techniques are employed which use Doppler ultrasonography to localize the
artery [
6].
COMPLICATIONS ! Arterial cannulation is a
relatively safe procedure, but is not without complications. Regardless of
site, the most common complications of arterial lines are:
• Bleeding
• Distal limb ischemia from thrombosis or
thromboembolism
• Air embolism
• Catheter-related infection
• Pain and swelling
• Pseudoaneurysm
• Arteriovenous fistula
Additional site specific complications are listed in Table 2 (
show
table 2).
Thrombosis ! Thrombosis can be detected by
Doppler ultrasonography in 5 to 25 percent of patients following arterial line
insertion, but clinically significant thrombosis occurs in less than 1 percent
[
7].
The incidence of thrombosis increases with longer duration of cannulation and
larger catheter diameters, and is more common in radial and dorsalis pedis
arteries because of their smaller size [
8].
Continuous or intermittent
heparin
flush systems have significantly reduced the incidence of catheter-related
thrombosis [
9].
Continuous flush systems deliver 3 to 15 units of
heparin
in approximately 3 mL per hour from an infusion bag pressurized to 300 mmHg. A
placebo-controlled, randomized trial of 40 ICU patients reported that a 1.4
percent solution of
sodium
citrate is as effective and safe as
heparin,
and may be useful when
heparin-associated
thrombocytopenia is suspected [
10].
(
See
"Clinical use of heparin and low molecular weight heparin").
Saline solutions are used to maintain patency in some centers, but may result
in higher catheter failure rates when compared with
heparin
[
11].
Infection ! Catheter-related infections are
discussed in depth separately. (
See
"Prevention of intravascular catheter-associated infections").
Reviewed briefly, there is a lower infection rate with arterial than venous
catheters, and colonized arterial catheters result less often in bacteremia
than colonized venous catheters. The reported incidence of arterial catheter
colonization varies from 5 to 10 percent, and appears similar for radial and
femoral lines [
12,13,14].
The risk for bacterial colonization of the catheter increases with the
duration of cannulation, rising to approximately 17 percent after six days [
15,16,17].
Nevertheless, scheduled changes of arterial lines have not been documented to
result in fewer infections or better patient outcomes. Guidewire exchange of
arterial catheters through the same insertion site has not been well studied
and is not recommended.
Controversy exists regarding the optimal interval at which arterial line
tubing and flush systems should be changed. While some authors recommend
scheduled replacement of such equipment every 24 to 96 hours to minimize the
opportunity for bacterial contamination and growth, others believe that this
practice is unnecessary or may be harmful. One study of 333 monitoring kits
and lines found that four out of six documented catheter infections occurred
within 48 hours after a scheduled flush bag change [
18].
The investigators hypothesized that changes of tubing may predispose to
infection because they require violations of the closed monitoring system. Our
approach is summarized in Table 3 (
show
table 3).
Air embolism ! Air embolism can occur when
the flush solution contains gas. Because the infusate is under a pressure of
300 mmHg, flushing may force air bubbles in a retrograde direction through the
radial, brachial, axillary, and subclavian arteries, where they can then
travel to the cerebral circulation via the vertebral arteries. In a primate
model, as little as 2 mL of air injected into the radial artery with a
standard pressurized infusion apparatus can result in significant cerebral air
emboli [
19].
Such emboli are more likely when patients are smaller and sitting upright. (
See
"Air Embolism").
Air introduced via arterial lines bypasses the pulmonary capillaries and thus
may cause more severe sequelae than air which is embolized through a
peripheral venous catheter. The latter occurs fairly commonly but rarely is
clinically apparent unless an anatomic right to left shunt is present. As an
example, one study performed electron-beam computed tomography of the chest on
208 patients following insertion of a peripheral venous catheter and detected
small air emboli in 5 percent, most commonly in the larger pulmonary arteries
[
20].
All patients were asymptomatic. (
See
"Electron beam (ultrafast) computed tomography for the evaluation of
cardiac disease and function").
Diagnostic blood loss ! Prior to obtaining
a blood specimen for laboratory analysis, 3 to 12 mL of blood typically is
wasted in order to avoid saline contamination. If frequent sampling is
required, this practice can result in substantial blood loss and necessitate
transfusion [
21,22].
The degree of diagnostic blood loss can be minimized by using a proximal
sampling port so that
heparin-free
blood can be obtained without wastage [
23].
Patients who require frequent arterial blood gases may also benefit from the
use of intraarterial blood gas monitoring. This is a new, fluorescent optode-based
technique which can perform arterial pH, PCO2, and PO2 measurements as
frequently as clinically required without violating the integrity of the
arterial catheter tubing system, or removing blood from the patient [
24].
QUALITY OF INVASIVE AND NONINVASIVE BLOOD
PRESSURE DATA ! Direct measurement by an arterial line is the gold
standard for determining blood pressure, provided the pressure transducer is
free of technical problems (see below). Values of mean arterial pressure
determined by direct measurement and sphygmomanometry generally correlate well
in healthy patients. (
See
"Technique of blood pressure measurement in the diagnosis of
hypertension").
Both manual and automated cuff pressures appear less reliable in
hemodynamically unstable patients. As an example, one study of 26
critically-ill patients found that indirect techniques underestimated systolic
pressure by an average of 17 mmHg, and overestimated diastolic pressure by 3
to 5 mmHg [
25].
Cuff pressures lose accuracy in the presence of shock, arrhythmias,
vasoconstrictor drugs, or calcified arteries [
26,27].
Noninvasive finger artery blood pressure monitoring correlates well with
radial artery blood pressure monitoring in various ambulatory settings [
28].
However, several small studies suggest that such devices have limited utility
when hemodynamic instability is present, and should not be relied upon in
settings where this is likely to occur [
29,30].
TECHNICAL SOURCES OF BLOOD PRESSURE MEASUREMENT
ERROR ! Despite the overall superiority of direct blood pressure
measurement methods, technical problems may result in systematic errors. The
major problems inherent to pressure monitoring with a catheter system are
dynamic response, zeroing drift, and transducer/monitor calibration [
31].
Dynamic response ! Dynamic response is
determined by two factors, the resonant frequency and the damping coefficient:
• The resonant or natural frequency of the system
is the frequency at which it oscillates when stimulated. Physiologic
peripheral arterial waveforms have a fundamental frequency of 3 to 5 Hz,
although some components may range up to 20 Hz [
32].
Thus, the resonant frequency of the system used to monitor arterial pressure
must be greater than 20 Hz to avoid ringing and systolic overshoot [
33]
• The damping coefficient is a measure of how
quickly an oscillating system comes to rest [
33].
A system with a high damping coefficient (eg, compliant tubing) absorbs
mechanical energy well and causes a diminution in the transmitted waveform.
The damping coefficient and resonant frequency of a monitoring system can be
assessed at the bedside by the fast-flush test, performed by briefly opening
and closing the valve in the continuous flush device (
show
figure 7). This produces a square wave displacement on the oscilloscope,
followed by ringing and a return to baseline. Connecting tubing with
stopcocks, excessive tubing lengths, and patient factors (such as tachycardia
and high output states) are common causes of underdamping. These factors
should be systematically addressed. On the other hand, air bubbles in the
tubing are a common source of overdamping; the bubbles can be cleared by
flushing the system through the stopcock.
Transducer position ! The transducer
position can be checked (the transducer can be "zeroed") by turning
the stopcock off to the patient and open to air, aligning the stopcock with
the level of the heart (usually approximated with the mid-axillary line), and
confirming that the monitor displays zero. This procedure should be repeated
when patient position is changed, when significant changes in blood pressure
are noted, and routinely every six to eight hours [
34].
Monitor and transducer calibration !
Routine calibration of the monitor and transducer is no longer necessary
because currently used, disposable transducers are standardized [
35].