Pulse oximetry
Hypoxemia constitutes an important cause of morbidity and mortality in the
intensive care unit (ICU), emergency department, and operating room. As a
result, accurate detection is important. It has long been recognized that a
patient's oxygenation is difficult to assess by physical examination alone.
Frank cyanosis does not develop until the deoxyhemoglobin level is 5 g/dL,
which corresponds to an arterial oxygen saturation (SaO2) around 67 percent [
1].
Furthermore, the threshold at which cyanosis becomes apparent is affected by
skin perfusion, skin pigmentation, and hemoglobin concentration [
2].
Blood gas analysis was for many years the accepted method of detecting
hypoxemia in critically ill patients, but this technique is painful, has
potential complications, and does not provide immediate or continuous data [
3].
(
See
"Measurement of arterial blood gases" and
see
"Placement and management of arterial catheters"). For these
reasons, pulse oximetry has become the standard for continuous and/or
noninvasive assessment of arterial oxygen saturation; it is now in such
ubiquitous use that it has been called the "fifth vital sign" [
4].
This card will review the theoretical and clinical aspects of pulse oximetry.
HISTORY ! Carl Matthes invented the first
noninvasive, nonpulsatile oximeter employing an ear probe in 1935. The device
used two wavelengths of light to compensate for variations in tissue thickness
and blood content. The development of oximetry intensified during World War
II, when a method was sought to monitor oxygenation in pilots flying at high
altitudes in pressurized cockpits [
3,5].
These early nonpulsatile devices did not measure true arterial saturation
because of interference from capillary and venous blood, and they were
unwieldy to use and transport [
6].
In 1970, scientists at Hewlett-Packard developed the first widely-used,
commercial ear oximeter that preferentially measured arterial saturation by
heating the tissue to 41ºC to increase blood flow [
5].
In 1974, Takuo Aoyagi discovered that arterial oxygen saturation could be
measured by looking for pulsations in the light signals coming through tissue.
This eliminated the need for heating the tissue, and his device is the
ancestor of the current generation of pulse oximeters [
6].
PRINCIPLES OF OPERATION ! The theoretical
basis for pulse oximetry derives from the Beer-Lambert law. This states that
the absorption of light of a given wavelength passing through a nonabsorbing
solvent which contains an absorbing solute is proportional to the product of
the solute concentration, the light path length, and an extinction
coefficient. The Beer-Lambert law can readily be applied to co-oximeters, in
which a sample of arterial blood can be placed within a cuvette and factors
such as light path length and solute concentration can be controlled [
7].
Application of this spectrophotometric principle to a noninvasive device,
where tissue is of varying thickness and blood flows in a pulsatile manner, is
more difficult. Modern pulse oximeters address these factors through the use
of two wavelengths of light and complex microprocessors. Deoxyhemoglobin
absorbs light maximally in the red band of the spectrum (600 to 750 nm), while
oxyhemoglobin absorbs maximally in the infrared band (850 to 1000 nm) [
8].
Absorbance at these two wavelengths is used to estimate saturation, which is
derived from the ratio of oxyhemoglobin to the sum of oxyhemoglobin plus
deoxyhemoglobin [
1].
Pulse oximeter probes consist of a photodetector and two light-emitting
diodes, one emitting at 660 nm and the other at 940 nm. The detector and
emitters are positioned facing each other through interposed tissue [
2].
Probes are most frequently placed on fingers or ear lobes [
9].
In infants, probes may also be placed on the palms, toes, feet, arms, cheeks,
tongue, penis, nose, or nasal septum [
10].
The photodiodes are switched on and off several hundred times per second, so
that light absorption by oxyhemoglobin and deoxyhemoglobin is recorded during
pulsatile and nonpulsatile flow [
11].
Absorption during pulsatile flow relates to the characteristics of arterial
blood plus background tissue and venous blood, whereas absorption during
nonpulsatile flow is due only to the background tissue and venous blood [
8,12]
(
show
figure 1). Absorption at the two wavelengths during pulsatile flow is
divided by absorption during nonpulsatile flow, and these ratios are fed into
an algorithm in the microprocessor to yield a saturation value. The displayed
value is an average based on the previous three to six seconds [
2,11].
In addition to SaO2, many pulse oximeters also display pulse rate and relative
pulse amplitude [
5].
The microprocessors of pulse oximeters are calibrated using reference tables
compiled by exposing healthy volunteers to decreasing FiO2 to yield SaO2
ranging from 100 to 75 percent by co-oximetry. Because it would be unethical
to intentionally generate lower saturations in volunteers, values for an SaO2
less than 75 percent are obtained by extrapolation from these volunteer data.
Pulse oximeter manufacturers claim that reported values between 70 and 100
percent are accurate to within \ 2 percent of the true value, while those
between 50 and 70 percent are within \ 3 percent. In practice, the cut-off
for acceptable accuracy is felt by many clinicians to be 80 percent (which
usually reflects a PaO2 of approximately 50 mmHg at a pH of 7.4), and varies
depending on the model of pulse oximeter used [
1,13,14].
The high accuracy of one type of pulse oximeter at SaO2 values ranging from 82
to 94 percent was confirmed in a study of 100 patients, although the the
accuracy of the instrument deteriorated at values outside these parameters [
15].
APPLICATIONS ! Pulse oximetry is indicated
in any clinical setting where hypoxemia may occur. These settings include
patient monitoring in emergency departments, operating rooms, postoperative
recovery areas, endoscopy suites, sleep and exercise laboratories, oral
surgery, cardiac catheterization, conscious sedation, labor and delivery, and
interfacility patient transfer [
2,16].
As an example of its benefits, pulse oximetry in a pediatric ICU decreases the
number of blood gases obtained and the duration of oxygen therapy without
jeopardizing patient outcome [
17].
In neonatal ICUs, oxygen toxicity in premature neonates can be limited by
titrating oxygen to an SaO2 of 90 percent (which usually reflects of PaO2 of
approximately 60 mmHg at a pH of 7.4) [
6,10].
(
See
"Oxygen toxicity"). However, just as pulse oximeters are not
ideally suited to detect severe hypoxemia, they are also not well suited to
detect hyperoxemia. Due to the shape of the O2-hemoglobin dissociation curve,
large changes in PaO2 may result in no change in O2 saturation if the
saturation is already near 100 percent. Therefore, in patients at risk for
profound hypoxemia or for hyperoxemia, SaO2 by pulse oximeter should be
verified by arterial blood gas analysis [
8].
Pulse oximetry offers the advantage of providing data on hemoglobin saturation
rather then PO2. SaO2 reflects the 98 percent of arterial oxygen content that
is normally carried by hemoglobin, while the PO2 directly measures only the
small amount of oxygen that is dissolved in plasma. PO2 commonly is used to
estimate arterial oxyhemoglobin saturation and oxygen content because the
dissolved and hemoglobin-bound oxygen pools are in equilibrium, but changes in
pH, temperature, and 2,3-diphosphoglycerate concentration alter the PO2-SaO2
relationship and may result in misleading calculations of oxyhemoglobin
saturation.
LIMITATIONS ! While pulse oximetry is a
convenient way of measuring arterial oxygenation, it does not assess
ventilation. Therefore, its use should be supplemented with arterial blood gas
analysis when hypoventilation is a concern [
16].
As an example, one inadvertently hypoventilated patient who was administered
100 percent oxygen during hip arthroplasty and monitored with pulse oximetry
alone developed a PaCO2 of 265 mmHg and an arterial pH of 6.65 despite
maintenance of oxygen saturations of 94 to 96 percent [
18].
In addition, because it does not measure PaO2, overreliance on pulse oximetry
may delay detection of clinically significant hypoxemia. A large decrease in
PO2 will not produce a significant fall in SaO2 until the steeper portion of
the oxygen hemoglobin dissociation curve is encountered at a PO2 of
approximately 60 to 70 mmHg. This is particularly important in patients
receiving supplemental oxygen. As an example, a fall in PaO2 in such a patient
from 140 to 65 mmHg would be required before a significant decrease in oxygen
saturation is detected.
Pulse oximetry results are signal-averaged over several seconds. Therefore,
the pulse oximeter may not detect a hypoxemic event until well after it has
occurred [
6].
This delay may be of particular significance when the device is being used for
monitoring during intubation.
TECHNICAL SOURCES OF ARTIFACT !
Interpretation of pulse oximetry readings must account for a variety of
factors which may artifactually influence the results. The best defense
against these potential sources of error is a high index of suspicion. If a
saturation reading is in doubt, a quick quality assurance test can be done by
the health care worker putting the probe on his or her own finger as near as
possible to the original patient site [
12].
The following are some of the more common potential technical sources of
error.
Ambient light ! Intense daylight,
fluorescent, incandescent, xenon, and infrared light sources have resulted in
spurious pulse oximetry readings [
2].
In such cases, the oximeter will often give a falsely low reading of 85
percent, the saturation at which the ratio of red to infrared light is 1.
Falsely elevated readings due to ambient light of normal intensity have been
reported, but are rare [
8].
Improper probe placement ! Due to partial
detachment of the probe, light from only one of the two light-emitting diodes
may pass through the tissue, causing either a falsely elevated or depressed
reading [
8].
A similar problem may occur in infants and small children, because the small
size of fingers or other tissues may result in differences in the path length
of one light source compared to that of the other. These problems can be
minimized by ensuring that the probe is properly attached with the light
sources and detectors opposite each other in a nontangential path [
19].
Placement of the sensor on the same extremity as a blood pressure cuff or
arterial line can cause erroneous readings and should be avoided [
20].
The choice of probe site may also affect accuracy; finger probes appear more
accurate than forehead, nose, or earlobe probes during low perfusion states [
9].
Motion artifact ! A poor signal-to-noise
ratio will cause signal artifact [
1,5].
This most commonly results from motion due to shivering, seizure activity,
pressure on the sensor, or transport of the patient by ambulance or
helicopter.
Electromagnetic radiation ! Radio frequency
emissions from magnetic resonance imaging (MRI) scanners may interfere with
pulse oximetry. In addition, second- and third-degree burns beneath pulse
oximeter probes have been reported in patients undergoing MRI studies [
10].
This complication is believed to result from the generation of electrical skin
currents beneath the looped pulse oximeter cables which act as an antenna.
Other sources of electromagnetic radiation, such as cellular phones and
electrocautery devices, can also interfere with pulse oximeters [
12,19].
PATIENT-RELATED SOURCES OF ERROR ! Pulse
oximetry readings may be inaccurate in certain clinical situations even if the
device is functioning properly and is free from external interference. The
results of pulse oximetry should be interpreted cautiously in the presence of
abnormal hemoglobins, nail polish, deeply pigmented skin, hypoperfusion,
anemia, venous congestion, or when certain dyes are in use.
Abnormal hemoglobins ! Abnormal
hemoglobins or hemoglobin variants may interfere with pulse oximetry if their
absorption properties are similar to those of oxyhemoglobin or deoxyhemoglobin.
• Carboxyhemoglobin absorbs approximately the
same amount of 660 nm light as does oxyhemoglobin; the pulse oximetry reading
therefore represents an inexact summation of oxyhemoglobin and
carboxyhemoglobin [
1,8,9,21,22]
(
show
figure 2). In cases of carbon monoxide poisoning or in chronic, heavy
smokers, a falsely reassuring pulse oximetry reading may mask life-threatening
arterial desaturation.
When carboxyhemoglobinemia is suspected, co-oximetry is required to accurately
measure oxyhemoglobin. Co-oximeters, which use four rather than two
wavelengths of light, detect oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin,
and methemoglobin, but require a sample of arterial whole blood [
6,17].
(
See
"Smoke inhalation").
• Methemoglobin absorbs at both 660 and 940 nm [
9].
Up to a methemoglobin level of 20 percent, SaO2 drops by about one-half of the
methemoglobin percentage. At higher methemoglobin levels, SaO2 tends toward 85
percent regardless of the true percentage of oxyhemoglobin [
1,10,23,24]
(
show
figure 3).
• Sickle cell hemoglobin generally produces pulse
oximeter readings similar to normal hemoglobin, but cases of falsely elevated
and falsely low readings have been reported [
16,25].
• Fetal hemoglobin gives pulse oximetry readings
clinically indistinguishable from those of adult hemoglobin [
5].
Nail polish ! The use of nail polish can
potentially affect pulse oximeter readings if the polish absorbs light at 660
nm or 940 nm [
1].
A small study of volunteers wearing black, green, and blue nail polish
revealed a drop in SaO2 of 3 percent, 5 percent, and 6 percent, respectively [
12].
This problem can be avoided by mounting the probe on the finger sideways,
rather than in a dorsal-ventral orientation [
10].
Red nail polish does not appear to have an effect on pulse oximetry readings.
Skin pigmentation ! Data on the effect of
skin pigmentation on pulse oximetry are conflicting [
12].
In theory, skin pigmentation should have no effect, since it should absorb at
a constant level and be subtracted out as part of the background in the SaO2
calculation; this hypothesis appears to apply to altered pigmentation due to
hyperbilirubinemia [
8].
However, readings which are erroneously elevated by approximately 4 percent
and a higher incidence of signal detection errors have been described in
African-American patients [
10].
Hypoperfusion ! Pulse oximetry readings
can be falsely low due to signal failure in the setting of hemodynamic
instability or poor limb perfusion from extremity elevation or
vasoconstriction. Measures which may improve the signal in these settings
include vigorous rubbing of the affected extremity, application of heat,
administration of a digital block to the finger being used, or the use of
topical vasodilators such as
nitroglycerin
paste or oil of wintergreen [
2,3,16].
Anemia ! In vitro and animal studies
suggest that pulse oximetry readings may be affected by hemoglobin
concentration [
16].
In vivo, low hemoglobin concentrations appears to cause falsely low readings
when the SaO2 is less than 80 percent [
10].
However, this effect is not clinically significant until the hemoglobin level
is less than 5 g/dL [
26].
Venous congestion ! Venous congestion due
to tricuspid valve incompetence or cardiomyopathy may yield falsely low SaO2
readings due to generation of venous pulsations. This results from the
instrument treating less oxygenated, pulsatile venous blood as part of the
arterial sample, thereby producing a smaller calculated SaO2 [
10].
Vital dyes ! The vital dyes
methylene
blue,
indocyanine
green, fluorescein, and indigo carmine can cause erroneously low pulse
oximetry readings.
Methylene
blue has the greatest impact, as it absorbs significantly at 670 nm.
However, these effects tend to be transient and resolve as the dyes are
diluted and metabolized [
5,12].